Efficacy of coronary stenting versus balloon angioplasty in small coronary arteries. Stent Restenosis Study (STRESS) Investigators

J Am Coll Cardiol. 1998 Feb;31(2):307-11. doi: 10.1016/s0735-1097(97)00511-1.

Abstract

Objectives: The goal of this study was to compare the efficacy of elective stent implantation and balloon angioplasty for new lesions in small coronary arteries.

Background: Palmaz-Schatz stents have been designed and approved by the Food and Drug Administration for use in coronary arteries with diameters > or = 3.0 mm. The efficacy of elective stent placement in smaller vessels has not been determined.

Methods: By quantitative coronary angiography, 331 patients in the Stent Restenosis Study (STRESS) I-II were determined to have a reference vessel < 3.0 mm in diameter. Of these, 163 patients were randomly assigned to stenting (mean diameter 2.69 +/- 0.21 mm), and 168 patients were assigned to angioplasty (mean diameter 2.64 +/- 0.24 mm). The primary end point was restenosis, defined as > or = 50% diameter stenosis at 6-month follow-up angiography. Clinical event rates at 1 year were assessed.

Results: Baseline clinical and angiographic characteristics were similar in the two groups. Procedural success was achieved in 100% of patients assigned to stenting and in 92% of patients assigned to angioplasty (p < 0.001). Abrupt closure within 30 days occurred in 3.6% of patients in both groups. Compared with angioplasty, stenting conferred a significantly larger postprocedural lumen diameter (2.26 vs. 1.80 mm, p < 0.001) and a larger lumen at 6 months (1.54 vs. 1.27 mm, p < 0.001). Restenosis (> or = 50% diameter stenosis at follow-up) occurred in 34% of patients assigned to stenting and in 55% of patients assigned to angioplasty (p < 0.001). At 1 year, event-free survival was achieved in 78% of the stent group and in 67% of the angioplasty group (p = 0.019).

Conclusions: These findings suggest that elective stent placement provides superior angiographic and clinical outcomes than balloon angioplasty in vessels slightly smaller than 3 mm.

Publication types

  • Clinical Trial
  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Angioplasty, Balloon, Coronary*
  • Chi-Square Distribution
  • Cohort Studies
  • Coronary Angiography
  • Coronary Disease / diagnostic imaging
  • Coronary Disease / pathology
  • Coronary Disease / therapy*
  • Coronary Vessels / pathology*
  • Disease-Free Survival
  • Elective Surgical Procedures
  • Equipment Design
  • Female
  • Follow-Up Studies
  • Humans
  • Linear Models
  • Male
  • Middle Aged
  • Recurrence
  • Stents*
  • Survival Rate
  • Treatment Outcome
  • United States
  • United States Food and Drug Administration