Diagnostic value of a new sensitive membrane based technique for instantaneous D-dimer evaluation in patients with clinically suspected deep venous thrombosis

Thromb Haemost. 1997 Apr;77(4):637-40.

Abstract

Background: Plasma D-Dimer analysis, using ELISA assays, has demonstrated in previous studies a high sensitivity, suggesting its utility in excluding deep venous thrombosis (DVT).

Aim: To assess the performance of a new rapid plasma D-Dimer ELISA measurement in suspected DVT patients with recent clinical signs, not exceeding one week.

Methods: A prospective study of patients admitted for a suspected recent DVT. Contrast venography or compression ultrasonography were performed within 24 h of admission. A new membrane based ELISA technique, which uses an immunofiltration and two complementary monoclonal antibodies was tested. Results were expressed as positive or negative. A standard plasma D-Dimer ELISA measurement was also performed. D-Dimer performances were assessed at the end of the study.

Results: 265/448 patients had a proven DVT (72 distal, 193 proximal). The sensitivity of the instantaneous method in the diagnosis of overall DVT is 92 +/- 3.4% (95% CI), and specificity is 36.6 +/- 6.9%. Positive predictive value is 67.7 +/- 4.8% and negative predictive value is 76.1 +/- 8.9%. Sensitivity and negative predictive values reach 97.9 and 94.3% in the diagnosis of proximal DVT, but only 76.3 and 79.7% in the diagnosis of distal DVT. Similar results are observed with the standard ELISA method.

Conclusion: This new rapid plasma D-Dimer measurement appears highly sensitive, and could substitute the older ELISA methods. Both methods provide lower sensitivity in the case of a distal DVT location.

Publication types

  • Clinical Trial
  • Controlled Clinical Trial

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Enzyme-Linked Immunosorbent Assay*
  • Evaluation Studies as Topic
  • Female
  • Fibrin Fibrinogen Degradation Products / analysis*
  • Humans
  • Male
  • Membranes, Artificial*
  • Middle Aged
  • Phlebography
  • Predictive Value of Tests
  • Prospective Studies
  • Reagent Kits, Diagnostic
  • Sensitivity and Specificity
  • Thrombophlebitis / blood
  • Thrombophlebitis / diagnosis*
  • Thrombophlebitis / diagnostic imaging
  • Time Factors
  • Ultrasonography

Substances

  • Fibrin Fibrinogen Degradation Products
  • Membranes, Artificial
  • Reagent Kits, Diagnostic
  • fibrin fragment D