Wide-mass-range axialization for high-resolution Fourier-transform ion cyclotron resonance mass spectrometry of externally generated ions

Rapid Commun Mass Spectrom. 1994 Aug;8(8):615-20. doi: 10.1002/rcm.1290080809.

Abstract

Ions generated in both continuous and pulsed-mode external ion sources are injected into an open-ended cylindrical Fourier-transform ion cyclotron resonance ion trap by means of an electrostatic ion guide consisting of concentric cylinder and wire electrodes to which a potential difference is applied. Broadband axialization during ion injection improves ion trapping efficiency by reducing the ion off-axis displacement. Significant improvements in both signal-to-noise ratio and mass resolving power, along with high mass selectivity are demonstrated for ions generated by external Cs+ secondary ion mass spectrometry or by matrix-assisted laser desorption/ionization.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, U.S. Gov't, Non-P.H.S.
  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Cesium / chemistry
  • Cyclotrons*
  • Electrodes
  • Lasers
  • Mass Spectrometry / instrumentation*
  • Spectroscopy, Fourier Transform Infrared