The aim of this study was to evaluate the validity and reproducibility between ozone measurements obtained with passive ozone monitors and those registered with a continuous ozone monitor, to determine the applicability of passive monitors in epidemiological research. The study was carried out during November and December 1992. Indoor and outdoor classroom air ozone concentrations were analyzed using 28 passive monitors and using a continuous monitor. The correlation between both measurements was highly significant (r = 0.089, p < 0.001), indicating a very good validity. Also, the correlation between the measurements obtained with two different passive monitors exposed concurrently was very high (r = 0.97, p < 0.001), indicating a good reproducibility in the measurements of the passive monitors. The relative error between the concentrations measured by the passive monitors and those from the continuous monitor tended to decrease with increasing ozone concentrations. The results suggest that passive monitors should be used to determine cumulative exposure of ozone exceeding 100 ppb, corresponding to an exposure period greater than five days, if used to analyze indoor air.