A Woven EndoBridge dislodging from an aneurysm 12 days after the implantation procedure: illustrative case

J Neurosurg Case Lessons. 2024 Oct 14;8(16):CASE24278. doi: 10.3171/CASE24278. Print 2024 Oct 14.

Abstract

Background: The Woven EndoBridge (WEB) is an effective embolization device. A few cases of WEB migration have been reported as adverse events, and most occurred during the procedure. A case of internal carotid artery (ICA) occlusion due to a trapped WEB, which occurred 12 days after the implantation procedure, is reported.

Observations: A woman in her 70s was referred for an enlarged left ICA aneurysm during follow-up. A WEB SL 5 × 3 was chosen as the embolic device, and the implantation procedure was completed uneventfully. The patient's postoperative course was good, and she was discharged from the hospital. However, 12 days later, she was brought to the authors' hospital because the WEB device was trapped and occluded in the left ICA. The trapped WEB could not be retrieved using a gooseneck snare or stent retriever, but suction from a Sofia catheter released the trapped device, allowing it to be repositioned within the aneurysm.

Lessons: Most cases of WEB migration occur during surgery. In the present case, migration occurred 12 days after surgery, and to the best of the authors' knowledge, no similar cases have been reported. It is important to keep in mind that there are cases similar to the current case. https://thejns.org/doi/10.3171/CASE24278.

Keywords: Woven EndoBridge; aneurysm embolization; dislodging.