Design of a diamond-based in-vessel soft x-ray detector for the SPARC tokamak

Rev Sci Instrum. 2024 Sep 1;95(9):093102. doi: 10.1063/5.0219522.

Abstract

The in-vessel silicon diode arrays that are used for soft x-ray detection in many tokamaks are sensitive to neutron damage, making them unsuitable for burning plasma devices such as SPARC. In such a device, the silicon diodes would need to be placed far from the plasma-limiting their field of view-or an alternative detector could be used. Here, we present the design of a camera containing an array of chemical vapor deposition single-crystal diamonds, which will be placed in the upper and lower port plugs of the SPARC tokamak with a large enough view of the poloidal cross section to enable tomographic inversion. The camera design presented here is optimized to provide a wide field of view of the poloidal cross section. Simulated plasma conditions are used to estimate the x-ray signal that this detector array will receive and to fine-tune the camera placement within the tokamak.