Evaluation of the analytical and clinical performance of two RT-PCR based point-of-care tests; Cepheid Xpert® Xpress CoV-2/Flu/RSV plus and SD BioSensor STANDARD™ M10 Flu/RSV/SARS-CoV-2

J Clin Virol. 2024 Jun:172:105674. doi: 10.1016/j.jcv.2024.105674. Epub 2024 Apr 16.

Abstract

Background: Rapid and accurate detection of viral respiratory infections is important for infection control measures. This study compares the analytical and clinical performance of the Xpert® Xpress CoV-2/Flu/RSV plus test ("Xpert", Cepheid) and the STANDARD™ M10 Flu/RSV/SARS-CoV-2 test ("M10", SD Biosensor). Both tests are quadruplex RT-PCR assays for rapid diagnosis of SARS-CoV-2, influenza A/B and RSV.

Study design: Analytical sensitivities were determined by limit of detection for SARS-CoV-2, influenza A, influenza B and RSV, respectively. Additionally, the clinical performance of the Xpert and the M10 tests was evaluated against standard-of-care RT-PCR by testing of 492 clinical specimens.

Results: The analytical sensitivities for Xpert versus M10 test was 10, 50, 50 and 300 versus 300, 200, 800 and 1500 copies/mL for SARS-CoV-2, influenza A, influenza B and RSV, respectively. Clinical sensitivity for the Xpert test was superior across all four pathogens compared to the M10 test. Xpert showed clinical sensitivity of 100 % in all Ct-ranges for all four pathogens whereas M10 showed clinical sensitivity of 100 % in the 25-30 Ct-range, 84-100 % in the 30-35 Ct-range and 47-67 % in the >35 Ct-range across the four pathogens. Translating into real-life clinical sensitivity, the Xpert would detect 100 % of all four pathogens, whereas M10 would detect 92.1, 92.4, 84.8 and 94.7 % for SARS-CoV-2, influenza A, influenza B and RSV.

Conclusion: This study demonstrates improved analytical and clinical performance of Xpert Xpress CoV-2/Flu/RSV plus compared to STANDARD M10 Flu/RSV/SARS-CoV-2, which is important for ensuring accuracy of diagnosis at all stages of a respiratory infection.

Keywords: GeneXpert; Influenza A/B; Molecular point-of-care test; RSV; SARS-CoV-2; STANDARD M10.

Publication types

  • Evaluation Study
  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • COVID-19 Nucleic Acid Testing / methods
  • COVID-19* / diagnosis
  • COVID-19* / virology
  • Humans
  • Influenza A virus* / genetics
  • Influenza A virus* / isolation & purification
  • Influenza B virus* / genetics
  • Influenza B virus* / isolation & purification
  • Influenza, Human* / diagnosis
  • Influenza, Human* / virology
  • Point-of-Care Testing
  • Respiratory Syncytial Virus Infections* / diagnosis
  • Respiratory Syncytial Virus Infections* / virology
  • Respiratory Syncytial Virus, Human / genetics
  • Respiratory Syncytial Virus, Human / isolation & purification
  • Reverse Transcriptase Polymerase Chain Reaction / methods
  • SARS-CoV-2* / genetics
  • SARS-CoV-2* / isolation & purification
  • Sensitivity and Specificity*