Introduction: Threaded conical centrifuge tubes are ubiquitous in biological laboratories and are frequently used for the storage/transport of potentially biohazardous samples. However, limited data are available on how frequently and from where these tubes leak. These data are valuable for laboratory biorisk management and to inform future studies on risks arising from the routine use of laboratory consumables.
Methods: The frequency of leaks from threaded conical centrifuge tubes was tested using a Glo Germ solution as a tracer. Conical tubes (15 and 50 mL) from several brands were filled, inverted, and placed on their side on the benchtop. After 1 h, the presence or absence of leaks on the benchtop surface, tube threads, and exterior was recorded.
Results: We observed that liquid leaked out of tubes that were apparently properly threaded in 2% of 15 mL tubes (confidence interval [95% CI] 1.4-2.6) and 1.4% of 50 mL tubes (95% CI 0.2-1.5). After opening, liquid was found on the threads on the outside of the tube in 20% of 15 mL tubes (95% CI 10-31) and 14% of 50 mL tubes (95% CI 1-28). We did not find sufficient evidence that differences in leak rates among brands were practically significant.
Conclusions: The fact that leaks were not uncommonly observed from conical centrifuge tubes suggests that mitigations for any hazard posed by a leak should be a component of every biorisk management strategy for protocols involving the manipulation of hazardous substances in these tubes. Further research should be conducted on other activities that could cause tubes to leak (such as centrifugation or vortexing) and should be completed to understand the risks associated with this consumable. Research into the costs and benefits of mitigating the risk of leaks from conical tubes is recommended.
Keywords: accident; biological risk management; conical centrifuge tube; laboratory procedures; laboratory safety; leak.