Background: The Neuroform Atlas Stent System is an established treatment modality for unruptured anterior and posterior circulation intracranial aneurysms. Location-specific results are needed to guide treatment decision-making. However, it is unclear whether there are differences in safety and efficacy outcomes between carotid and more distal anterior circulation aneurysms.
Methods: The ATLAS IDE trial was a prospective, multicenter, single-arm, open-label interventional study that evaluated the safety and efficacy of the Neuroform Atlas Stent System. We compared differences in efficacy and safety outcomes of proximal internal carotid artery (ICA) versus distal and bifurcation anterior circulation aneurysms.
Results: Of 182 cases, there were 70 aneurysms in the ICA and 112 in the distal anterior circulation (including ICA terminus/bifurcation). There were no significant differences in the primary efficacy endpoint (85.5% vs 83.9%, p=0.78) and complete aneurysm occlusion rates (88.7% vs 87.9%, p=0.78) between proximal ICA aneurysms and distal aneurysms, respectively. Complications were more often encountered in distal and bifurcation aneurysms, but the overall rate of major safety events was low and comparable between the two groups (1.4% vs 6.3%, p=0.14). Recanalization and retreatment rates were also similar between the groups.
Conclusion: The results of this study suggest that the Neuroform Atlas Stent System is a safe and efficacious treatment modality for unruptured anterior circulation intracranial aneurysms, regardless of aneurysm location.
Trial registration number: NCT02340585.
Keywords: Aneurysm; Coil; Device; Intervention; Stent.
© Author(s) (or their employer(s)) 2024. Re-use permitted under CC BY-NC. No commercial re-use. See rights and permissions. Published by BMJ.