The CINCH-FMR postmarket registry: Real-world long-term outcomes with percutaneous mitral valve repair with the Carillon Mitral Contour System®

Cardiovasc Revasc Med. 2024 Mar:60:35-40. doi: 10.1016/j.carrev.2023.09.007. Epub 2023 Sep 23.

Abstract

Background: The Carillon® Mitral Contour System® has been studied in 4 prospective controlled studies in the treatment of functional mitral regurgitation (FMR) where it has been found to reduce mitral regurgitation, reduce left ventricular and atrial volumes, and be associated with improvements in clinical parameters.

Aims: The CINCH post-market registry is designed to evaluate immediate, mid-term and long-term outcomes from a post-approval study of the Carillon® device evaluated in real-world practice.

Methods: The CINCH post-market registry is a single-arm study of percutaneous mitral annuloplasty with the Carillon device in patients with functional (secondary) mitral regurgitation and symptomatic congestive heart failure when utilized in real-world conditions. Patient selection, echocardiographic hemodynamic measurements, and patient follow-up requirements were performed per standard of care at each institution.

Results: A total of 101 patients treated with the Carillon device at 13 sites in Germany were enrolled in the CINCH registry. The mean age was 75 ± 9 years, 57 % were male, and patient presentation included primarily NYHA class III (69 %) with MR grade 3 (68 %). Over 5 years of follow-up, all-cause mortality was 40.1 %, the incidence of HFH was 53.9 %, and the composite outcome of HFH or death was 66.4 %. At each follow-up interval through 5 years, statistically significant reductions in NYHA class (p < 0.05) and MR grade (p < 0.01) were reported.

Conclusions: In this "real world" registry of the Carillon Mitral Contour System, procedural safety and medium-term follow-up outcomes is similar to the outcomes seen in the prospective, controlled clinical trials, despite being used in populations of patients that extend outside of those studied in the trials. The use of this therapy in patients with atrial functional mitral regurgitation, and heart failure with preserved ejection fraction, was notable, since these types of patients were excluded from the prospective, controlled trials. This supports possible additional patient populations who might benefit from this type of mechanical therapy. The safety profile of this therapy in this registry and in the earlier trials may support a potential role in earlier forms of secondary mitral regurgitation.

Keywords: CINCH; Carillon; Heart failure; Indirect annuloplasty; Mitral regurgitation.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Echocardiography
  • Female
  • Heart Failure*
  • Heart Valve Prosthesis Implantation* / adverse effects
  • Humans
  • Male
  • Mitral Valve / diagnostic imaging
  • Mitral Valve / surgery
  • Mitral Valve Annuloplasty* / adverse effects
  • Mitral Valve Insufficiency* / diagnostic imaging
  • Mitral Valve Insufficiency* / surgery
  • Prospective Studies
  • Registries
  • Treatment Outcome