Objectives: To evaluate the accuracy of implant placement using a dynamic navigation system in fully edentulous jaws and to analyze the influence of implant distribution on implant position accuracy.
Materials and methods: Edentulous patients who received implant placement using a dynamic navigation system were included. Four to six mini screws were placed in the edentulous jaw under local anesthesia as fiducial markers. Then patients received CBCT scans. Virtual implant positions were designed in the planning software based on CBCT data. Under local anesthesia, implants were inserted under the guidance of the dynamic navigation system. CBCTs were taken following implant placement. The deviation between the actual and planned implant positions was measured by comparing the pre- and postsurgery CBCT.
Results: A total of 13 edentulous patients with 13 edentulous maxillae and 7 edentulous mandibles were included, and 108 implants were placed. The average linear deviations at the implant entry point and apex were 1.08 ± 0.52 mm and 1.15 ± 0.60 mm, respectively. The average angular deviation was 2.85 ± 1.20°. No significant difference was detected in linear and angular deviations between the maxillary and mandibular implants, neither between the anterior and posterior implants.
Conclusions: The dynamic navigation system provides high accuracy for implant placement in fully edentulous jaws, while the distribution of the implants showed little impact on implant position accuracy.
Keywords: CAIS; accuracy; dynamic navigation; edentulous; implant.
© 2023 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.