Both single- and double-outflow cavopulmonary assist devices (CPADs) were recently proposed for the Fontan population, whereas single-outflow configurations were evaluated in large animal trials and double-outflow concepts are lacking an equivalent in vivo assessment. The aim of this study was to test the hemodynamic properties of a double-outflow CPAD device in an acute sheep model. The two inflow cannulae of a CPAD were anastomosed to the caval veins. Outflow graft connection was performed via end-to-side anastomosis to the right (RPA) and main pulmonary artery (MPA). Speed ramp protocols were conducted, and hemodynamic effects were monitored in terms of caval flows, cardiac output (CO), central venous pressure (CVP), pulmonary artery pressure (PAP), and left atrial pressure (LAP). Six experiments were conducted (53.35 ± 5.1 kg). In three experiments, the animal model was established, the CPAD was examined, and restoration of biventricular equivalency in terms of venous return was achieved. Venous pressures (CVP) declined linearly with increasing pump speed (r > 0.879), whereas caval flow (r > 0.973), CO (r > 0.993), PAP (r > 0.973), and LAP (r > 0.408) increased. Despite the considerable complexity of the sheep model caused by the sheep pulmonary arterial anatomy that requires substantial graft bending, the CPAD was evaluated in three acute experiments and showed the potential to completely substitute a subpulmonary ventricle.