Two-Year Outcomes With a Next-Generation Left Atrial Appendage Device: Final Results of the PINNACLE FLX Trial

J Am Heart Assoc. 2023 Feb 21;12(4):e026295. doi: 10.1161/JAHA.122.026295. Epub 2023 Feb 15.

Abstract

Background The PINNACLE FLX (Protection Against Embolism for Non-valvular AF [Atrial Fibrillation] Patients: Investigational Device Evaluation of the Watchman FLX LAA [Left Atrial Appendage] Closure Technology) trial evaluated the safety and efficacy of a next-generation left atrial appendage closure device (WATCHMAN FLX; Boston Scientific, Marlborough, MA). At 1 year, the study met the primary end points of safety and anatomical efficacy/appendage closure. This final report of the PINNACLE FLX trial includes the prespecified secondary end point of ischemic stroke or systemic embolism at 2 years, also making it the first report of 2-year outcomes with this next-generation left atrial appendage closure device. Methods and Results Patients with nonvalvular atrial fibrillation with CHA2DS2-VASc score ≥2 (men) or ≥3 (women), with an appropriate rationale for left atrial appendage closure, were enrolled to receive the left atrial appendage closure device at 29 US centers. Adverse events were assessed by an independent clinical events committee, and imaging was assessed by independent core laboratories. Among 395 implanted patients (36% women; mean age, 74 years; CHA2DS2-VASc, 4.2±1.5), the secondary efficacy end point of 2-year ischemic stroke or systemic embolism was met, with an absolute rate of 3.4% (annualized rate, 1.7%) and an upper 1-sided 95% confidence bound of 5.3%, which was superior to the 8.7% performance goal. Two-year rates of adverse events were as follows: 9.3% all-cause mortality, 5.5% cardiovascular death, 3.4% all stroke, and 10.1% major bleeding (Bleeding Academic Research Consortium 3 or 5). There were no additional systemic embolisms, device embolizations, pericardial effusions, or symptomatic device-related thrombi after 1 year. Conclusions The secondary end point of 2-year stroke or systemic embolism was met at 3.4%. In these final results of the PINNACLE FLX trial, the next-generation WATCHMAN FLX device demonstrated favorable safety and efficacy outcomes.

URL: https://www.clinicaltrials.gov; Unique identifier: NCT02702271.

Keywords: WATCHMAN FLX; atrial fibrillation; left atrial appendage closure.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Atrial Appendage* / diagnostic imaging
  • Atrial Appendage* / surgery
  • Atrial Fibrillation* / complications
  • Atrial Fibrillation* / surgery
  • Cardiac Catheterization / adverse effects
  • Embolism* / complications
  • Embolism* / prevention & control
  • Female
  • Hemorrhage / etiology
  • Humans
  • Ischemic Stroke* / etiology
  • Male
  • Stroke* / complications
  • Stroke* / prevention & control
  • Treatment Outcome