Volumetric absorptive microsampling-LC-MS/MS assays for quantitation of giredestrant in dried human whole blood

Bioanalysis. 2022 Nov;14(21):1377-1389. doi: 10.4155/bio-2022-0189. Epub 2023 Jan 19.

Abstract

Volumetric absorption microsampling devices offer minimally invasive and user-friendly collection of capillary blood in volumes as low as 10 μl. Herein we describe the assay validation for determination of the selective estrogen receptor degrader giredestrant (GDC-9545) in dried human whole blood collected using the Mitra® and Tasso-M20 devices. Both LC-MS/MS assays met validation acceptance criteria for the linear range 1-1000 ng/ml giredestrant. Mitra and Tasso-M20 samples were stable for 84 and 28 days at ambient conditions, respectively, and for 7-9 days at 40 and -70°C. Blood hematocrit, hyperlipidemia and anticoagulant did not impact quantitation of giredestrant. These validated assays are suitable for the determination of giredestrant in dried blood samples collected using Mitra and Tasso-M20 microsampling devices.

Keywords: LC–MS/MS; Mitra; Tasso-M20; VAMS; anticoagulant; hematocrit; microsampling; validation; whole blood.