In January 2022, our institution launched a comprehensive cancer genome profiling program on 10 cancer types using a non-IVD solution named the TruSight Oncology 500 Assay provided by Illumina®. The assay analyzes both DNA and RNA, identifying Single-Nucleotide Variants (SNV)s and Insertion-Deletion (InDel) in 523 genes, as well as known and unknown fusions and splicing variants in 55 genes and Copy Number Alterations (CNVs), Mutational Tumor Burden (MTB) and Microsatellite Instability (MSI). According to the current European IVD Directive 98/79/EC, an internal validation was performed before running the test. A dedicated open-source bioinformatics pipeline was developed for data postprocessing, panel assessment and embedding in high-performance computing framework using the container technology to ensure scalability and reproducibility. Our protocols, applied to 71 DNA and 64 RNA samples, showed full agreement between the TruSight Oncology 500 assay and standard approaches, with only minor limitations, allowing to routinely perform our protocol in patient screening.
Keywords: bioinformatics analysis; coverage analysis; next-generation sequencing; oncology; panel validation.