Background and purpose: During endoscopic ultrasound-guided hepaticogastrostomy (EUS-HGS), tract dilation is one of the most important steps, and the placement of conventional metal stents with 8.5 Fr delivery devices is difficult due to the large outer shape of the device. Fine-gauge balloon catheters have become popular because of their stricture penetration ability and ease of dilation. This study aimed to evaluate the utility of fine-gauge balloon catheters.
Patients and methods: This retrospective study involved 38 patients who underwent conventional metal stent placement. The patients were classified into two groups: those who underwent dilation with a fine-gauge balloon catheter before initial metal stenting (balloon dilation group) and those who underwent bougie dilation only (non-balloon dilation group). We evaluated the stenting success rate after initial dilation and adverse events.
Results: Seventeen and twenty-one patients were included in the balloon dilation and non-balloon dilation groups, respectively. The stenting success rate after initial dilation was 100% (17/17) in the balloon dilation group and 71.4% (15/21) in the non-balloon dilation group (p = 0.024). As adverse events, peritonitis was observed in one case (4.8%) in the balloon dilation group, and in three cases (14.3%) in the non-balloon dilation group (p = 0.613).
Conclusions: Dilation using a fine-gauge balloon catheter before conventional metal stent with 8.5 Fr delivery device placement is considered effective in EUS-HGS.
Keywords: EUS-guided biliary drainage; biliary obstruction; hepaticogastrostomy; interventional EUS.