Background and aims: Artificial intelligence-based computer-aid detection (CADe) devices have been recently tested in colonoscopies, increasing the adenoma detection rate (ADR), mainly in Asian populations. However, evidence for the benefit of these devices in the occidental population is still low. We tested a new CADe device, namely, ENDO-AID (OIP-1) (Olympus, Tokyo, Japan), in clinical practice.
Methods: This randomized controlled trial included 370 consecutive patients who were randomized 1:1 to CADe (n = 185) versus standard exploration (n = 185) from November 2021 to January 2022. The primary endpoint was the ADR. Advanced adenoma was defined as ≥10 mm, harboring high-grade dysplasia, or with a villous pattern. Otherwise, the adenoma was nonadvanced. ADR was assessed in both groups stratified by endoscopist ADR and colon cleansing.
Results: In the intention-to-treat analysis, the ADR was 55.1% (102/185) in the CADe group and 43.8% (81/185) in the control group (P = .029). Nonadvanced ADRs (54.8% vs 40.8%, P = .01) and flat ADRs (39.4 vs 24.8, P = .006), polyp detection rate (67.1% vs 51%; P = .004), and number of adenomas per colonoscopy were significantly higher in the CADe group than in the control group (median [25th-75th percentile], 1 [0-2] vs 0 [0-1.5], respectively; P = .014). No significant differences were found in serrated ADR. After stratification by endoscopist and bowel cleansing, no statistically significant differences in ADR were found.
Conclusions: Colonoscopy assisted by ENDO-AID (OIP-1) increases ADR and number of adenomas per colonoscopy, suggesting it may aid in the detection of colorectal neoplastic lesions, especially because of its detection of diminutive and flat adenomas. (Clinical trial registration number: NCT04945044.).
Copyright © 2023 American Society for Gastrointestinal Endoscopy. Published by Elsevier Inc. All rights reserved.