A Comparison of Revision Rates and Dislocation After Primary Total Hip Arthroplasty with 28, 32, and 36-mm Femoral Heads and Different Cup Sizes: An Analysis of 188,591 Primary Total Hip Arthroplasties

J Bone Joint Surg Am. 2022 Aug 17;104(16):1462-1474. doi: 10.2106/JBJS.21.01101. Epub 2022 Jun 16.

Abstract

This article was updated on August 17, 2022, because of previous errors, which were discovered after the preliminary version of the article was posted online. On page 1462, in the first sentence of the Abstract section entitled "Results," the phrase that had read "and 36-mm heads had fewer dislocations than 28-mm (HR = 0.33 [95% CI, 0.16 to 0.68]; p = 0.003), but more dislocations than 32-mm heads (HR for >2 weeks = 2.25 [95% CI, 1.13 to 4.49]; p = 0.021)" now reads "and 36-mm heads had fewer dislocations than 28-mm (HR = 0.33 [95% CI, 0.16 to 0.68]; p = 0.003) and 32-mm heads (HR for ≥2 weeks = 0.44 [95% CI, 0.22 to 0.88]; p = 0.021)." On page 1468, in the last sentence of the section entitled "Acetabular Components with a Diameter of <51 mm," the phrase that had read "and HR for ≥2 weeks = 2.25 [95% CI, 1.13 to 4.49; p = 0.021]) ( Fig. 3 )" now reads "and HR for ≥2 weeks = 0.44 [95% CI, 0.22 to 0.88; p = 0.021]) ( Fig. 3 )." Finally, on page 1466, in the upper right corner of Figure 3 , under "32mm vs 36mm," the second line that had read "2Wks+: HR=2.25 (1.13, 4.49), p=0.021" now reads "2Wks+: HR=0.44 (0.22, 0.88), p= 0.021."

Plain language summary

The acetabular component diameter can influence the choice of femoral head size in total hip arthroplasty (THA). We compared the rates of revision by femoral head size for different acetabular component sizes. Data from the Australian Orthopaedic Association National Joint Replacement Registry were analyzed for patients undergoing primary THA for a diagnosis of osteoarthritis from September 1999 to December 2019. Acetabular components were stratified into quartiles by size: <51 mm, 51 to 53 mm, 54 to 55 mm, and 56 to 66 mm. Femoral head sizes of 28 mm, 32 mm, and 36 mm were compared for each cup size. The primary outcome was the cumulative percent revision (CPR) for all aseptic causes and for dislocation. The results were adjusted for age, sex, femoral fixation, femoral head material, year of surgery, and surgical approach and were stratified by femoral head material. For acetabular components of <51 mm, 32-mm (hazard ratio [HR] = 0.75 [95% confidence interval (CI), 0.57 to 0.97]; p = 0.031) and 36-mm femoral heads (HR = 0.58 [95% CI, 0.38 to 0.87]; p = 0.008) had a lower CPR for aseptic causes than 28-mm heads; and 36-mm heads had fewer dislocations than 28-mm (HR = 0.33 [95% CI, 0.16 to 0.68]; p = 0.003), and 32-mm heads (HR for ≥2 weeks = 0.44 [95% CI, 0.22 to 0.88]; p = 0.021). For 51 to 53-mm, 54 to 55-mm, and 56 to 66-mm-diameter acetabular components, there was no difference in the CPR for aseptic causes among head sizes. A femoral head size of 36 mm had fewer dislocations in the first 2 weeks than a 32-mm head for the 51 to 53-mm acetabular components (HR for <2 weeks = 3.79 [95% CI, 1.23 to 11.67]; p = 0.020) and for the entire period for 56 to 66-mm acetabular components (HR = 1.53 [95% CI, 1.05 to 2.23]; p = 0.028). The reasons for revision differed for each femoral head size. There was no difference in the CPR between metal and ceramic heads. There is no clear advantage to any single head size except with acetabular components of <51 mm, in which 32-mm and 36-mm femoral heads had lower rates of aseptic revision. If stability is prioritized, 36-mm femoral heads may be indicated. Therapeutic Level III . See Instructions for Authors for a complete description of levels of evidence.

MeSH terms

  • Arthroplasty, Replacement, Hip* / adverse effects
  • Femur Head / surgery
  • Hip Dislocation* / complications
  • Hip Dislocation* / surgery
  • Hip Prosthesis* / adverse effects
  • Humans
  • Joint Dislocations* / complications
  • Prosthesis Design
  • Prosthesis Failure
  • Registries
  • Reoperation / adverse effects
  • Risk Factors