Diagnostic accuracy of AS-OCT vs gonioscopy for detecting angle closure: a systematic review and meta-analysis

Graefes Arch Clin Exp Ophthalmol. 2022 Jan;260(1):1-23. doi: 10.1007/s00417-021-05271-4. Epub 2021 Jul 5.

Abstract

Purpose: This study aims to review the literature that compares the accuracy of Anterior Segment-Optical Coherence Tomography (AS-OCT) against gonioscopy in detecting eyes with angle closure. It is currently unclear how AS-OCT fits into clinical practice for detecting angle closure. This is a systematic review and meta-analysis.

Methods: A literature search was performed on Medline, Embase, Scopus and the Cochrane Central Register of Controlled Trials to identify studies that investigated the diagnostic accuracy of AS-OCT in detecting eyes with angle closure as diagnosed by gonioscopy. Eligible studies included in the analysis met stringent inclusion criteria determining the sensitivity and specificity of AS-OCT.

Results: The initial search identified 727 studies, of which 23 were included in the final analysis. We found substantial variation in the parameters being studied and methodologies. The sensitivity of AS-OCT ranged from 46 to 100% (median 87%). Twenty-one studies identified parameters that showed sensitivity above 80%. The specificity ranged from 55.3 to 100% (median 84%).

Conclusion: AS-OCT demonstrates good sensitivity for detecting angle closure. It may provide an avenue to address high rates of undiagnosed angle closure, such as found in developing Asian countries. However, AS-OCT is not yet able to replace gonioscopy. Clinicians should consider whether the diagnostic accuracy of AS-OCT is acceptable for their specific clinical use before adopting it. More studies are needed to determine the utility of AS-OCT, including longitudinal studies to determine the significance of eyes classified to have closed angles by AS-OCT but open on gonioscopy.

Keywords: Anterior segment optical coherence tomography; Glaucoma; Gonioscopy; Narrow angle; OCT; Screening.

Publication types

  • Meta-Analysis
  • Review
  • Systematic Review

MeSH terms

  • Anterior Eye Segment / diagnostic imaging
  • Glaucoma, Angle-Closure* / diagnosis
  • Gonioscopy
  • Humans
  • Intraocular Pressure
  • Tomography, Optical Coherence*