Reconstruction of infected post-traumatic bone defects of the distal femur with the Compress implant. Preliminary results of a staged non-biological strategy

Injury. 2021 Mar;52(3):606-615. doi: 10.1016/j.injury.2020.10.016. Epub 2020 Oct 6.

Abstract

Introduction: Infected post-traumatic distal femur defects remain a therapeutic challenge. Non-biological reconstruction offers an option for avoiding complex biological knee arthrodesis procedures. The Compress implant is an alternative to the traditional distal femur stemmed megaprosthesis. The aim of this study is to analyse the first patients treated with a distal femur Compress prosthesis to manage massive infected post-traumatic defects of the distal femur with joint involvement.

Methods: We retrospectively reviewed all patients with massive infected defects of the distal femur where this implant was used in a two-stage strategy, together with an antibacterial coating hydrogel (DAC). The specific protocol, microbiological data, clinical and radiological results, complications, functional results and prosthesis survivorship were determined. Follow-up was for a minimum of 12 months, or until implant removal.

Results: Ten patients (11 Compress implants) with a mean age of 52 years (range 35-73) were included. On average, patients had undergone 4.4 previous surgical procedures before index surgery. The mean bone defect was 14 cm (range 8-21). After a median follow-up of 27 months (range 12-50 months) no patient had presented with recurrence of the infection, and limb salvage was achieved in all cases. Two patients suffered aseptic loosening which required revision of the femoral component. The short-term survivorship of the implant in our series was 81.8% at 4 years, with all failures occurring in the first 7 months. After this 7-month time threshold, we encountered no further loosening. Regarding functional outcomes, patients had a mean knee ROM of -4/86, expressed high overall satisfaction with the procedure according to the SAPS scale, and had an average LEFS of 52.5% (40-72.5%).

Conclusion: Non-biological reconstruction of the distal femur with the Compress implant is a valid option in selected patients with massive infected defects with joint involvement. Survivorship was high, with all loosening occurring in the first months after surgery-representing a failure in the osseointegration of the implant.

Keywords: CompressⓇ; distal femur fractures; infected bone defects; megaprosthesis; post-traumatic bone defects.

MeSH terms

  • Adult
  • Aged
  • Femur* / diagnostic imaging
  • Femur* / surgery
  • Humans
  • Middle Aged
  • Prosthesis Design
  • Prosthesis Failure
  • Prosthesis Implantation*
  • Reoperation
  • Retrospective Studies
  • Treatment Outcome