Background: Currently, 2 third-generation transcatheter valves, 29-mm Sapien-3 and 34-mm Evolut-R (ER), are indicated for large sized aortic annuli. We analyzed short and 1-year performance of these valves in patients with large (area ≥575 mm2 or perimeter ≥85 mm) and extra-large (≥683 mm2 or ≥94.2 mm) aortic annuli undergoing transcatheter aortic valve replacement.
Methods: A total of 833 patients across 12 centers with symptomatic aortic stenosis and large aortic annuli underwent transcatheter aortic valve replacement with 29-mm Sapien-3 (n=640) or 34-mm ER (n=193). Clinical, anatomic, and procedural characteristics were collected, and Valve Academic Research Consortium-2 outcomes were reported.
Results: Median aortic annulus area and perimeter were 617 mm2 (591-657) and 89.1 mm (87.0-92.1), respectively (704 mm2 [689-743] and 96.0 mm [94.5-97.9] in the subgroup of 124 patients with extra-large annuli). Overall device success was 94.3% (Sapien-3, 95.8% and ER, 89.3%; P=0.001), with a higher rate of significant paravalvular leak (P=0.004), second valve implantation (P=0.013), and valve embolization (P=0.009) in the ER group. Thirty-day and 1-year mortality was 2.4% and 9.2%, respectively, without differences between groups. Valve hemodynamics were excellent (mean gradient, 8.8±3.6 mm Hg; 3.3% rate of moderate-severe paravalvular leak) in the extra-large annulus, without differences compared with the large annulus group.
Conclusions: In patients with large and extra-large aortic annuli, transcatheter aortic valve replacement using 29-mm Sapien-3 and 34-mm ER is safe and feasible. Observed differences in clinical outcomes and hemodynamic performance may guide valve choice in this cohort of patients undergoing transcatheter aortic valve replacement.
Keywords: aortic valve stenosis; cohort studies; hemodynamics; humans; transcatheter aortic valve replacement.