Abstract

PRéCIS:: Averaging triplicate en face angiograms of the radial peripapillary capillary (RPC) plexus with optical coherence tomography angiography (OCTA) improves vessel visualization, reduces vessel density parameters, and increases the diagnostic accuracy for glaucoma of one such parameter.

Purpose: The purpose of this study was to test the hypothesis that triplicate averaging of the RPC layer improves visualization and diagnostic accuracy of OCTA for glaucoma.

Materials and methods: This is a cross-sectional study involving 63 primary open-angle glaucoma patients and 70 age-matched glaucoma suspects. Triplicate 6×6 mm OCTA scans of the optic nerve head were acquired, and the RPC layer was extracted. RPC en face images were registered and averaged. Parameters of global entropy, global standard deviation, local texture correlation, local homogeneity, signal-to-noise ratio, and intercapillary distance were used to measure the change in visualization with averaging. Vessel area density (VAD), vessel skeleton density (VSD), and flux parameters were calculated in a 2.8 mm annulus excluding the optic disc. The diagnostic accuracy of these parameters for glaucoma was assessed by calculating the area under the receiver operating curve (AUC) values.

Results: Three-frame averaging resulted in decreased global entropy and global standard deviation (Ps<0.001), and increased local texture correlation, local homogeneity, signal-to-noise ratio, and intercapillary distance (Ps<0.001). Averaged images also had reduced VAD, VSD, and flux (Ps<0.001). AUC was significantly increased for VSD after image averaging (P=0.018), while no significant change in AUC was observed for VAD (P=0.229) or flux (P=0.193).

Conclusions: Triplicate averaging improves visualization of the RPC layer and the diagnostic accuracy of VSD for glaucoma. The impact of image averaging on OCTA diagnostic performance and other potential applications warrants further exploration.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Area Under Curve
  • Cross-Sectional Studies
  • Female
  • Fluorescein Angiography*
  • Glaucoma, Open-Angle / diagnostic imaging*
  • Glaucoma, Open-Angle / physiopathology
  • Healthy Volunteers
  • Humans
  • Intraocular Pressure
  • Male
  • Middle Aged
  • Ocular Hypertension / diagnostic imaging
  • Ocular Hypertension / physiopathology
  • Optic Disk / blood supply
  • Prospective Studies
  • Retinal Vessels / diagnostic imaging*
  • Retinal Vessels / physiopathology
  • Tomography, Optical Coherence*