5-Year Outcomes After TAVR With Balloon-Expandable Versus Self-Expanding Valves: Results From the CHOICE Randomized Clinical Trial

JACC Cardiovasc Interv. 2020 May 11;13(9):1071-1082. doi: 10.1016/j.jcin.2019.12.026. Epub 2020 Apr 15.

Abstract

Objectives: The purpose of this study was to evaluate clinical and echocardiographic outcome data of the CHOICE (Randomized Comparison of Transcatheter Heart Valves in High Risk Patients with Severe Aortic Stenosis: Medtronic CoreValve Versus Edwards SAPIEN XT) trial at 5 years.

Background: The CHOICE trial was designed to compare device performance of a balloon-expandable (BE) transcatheter heart valve (THV) versus a self-expanding (SE) THV.

Methods: The CHOICE trial is an investigator-initiated trial that randomized 241 high-risk patients with severe symptomatic aortic stenosis and an anatomy suitable for treatment with both BE and SE THVs to transfemoral transcatheter aortic valve replacement with either device. The primary endpoint was device success. Patients were followed up to 5 years, with assessment of clinical outcomes, and echocardiographic evaluation of valve function and THV durability.

Results: After 5 years, there were no statistically significant differences between BE and SE valves in the cumulative incidence of death from any cause (53.4% vs. 47.6%; p = 0.38), death from cardiovascular causes (31.6% vs. 21.5%; p = 0.12), all strokes (17.5% vs. 16.5%; p = 0.73), and repeat hospitalization for heart failure (28.9% vs. 22.5%; p = 0.75). SE patients had larger prosthetic valve area (1.6 ± 0.5 cm2 vs. 1.9 ± 0.5 cm2; p = 0.02) with a lower mean transprosthetic gradient (12.2 ± 8.7 mm Hg vs. 6.9 ± 2.7 mm Hg; p = 0.001) at 5 years. No differences were observed in the rates of paravalvular regurgitation. Clinical valve thrombosis occurred in 7 BE patients (7.3%) and 1 SE patient (0.8%; p = 0.06), and moderate or severe structural valve deterioration in 6 BE patients (6.6%) and no SE patient (0%; p = 0.018). The rate of bioprosthetic valve failure was low and not significantly different between both groups (4.1% vs. 3.4%; p = 0.63).

Conclusions: Five-year follow-up of patients in the CHOICE trial revealed clinical outcomes after transfemoral transcatheter aortic valve replacement with early-generation BE and SE valves that were not statistically significantly different, with limited statistical power. Forward flow hemodynamics were significantly better with the SE valve. Moderate or severe structural valve deterioration was uncommon but occurred more frequently with the BE valve. (A Comparison of Transcatheter Heart Valves in High Risk Patients With Severe Aortic Stenosis: The CHOICE Trial [CHOICE]; NCT01645202).

Keywords: TAVR; aortic stenosis; balloon-expandable; durability; self-expanding.

Publication types

  • Comparative Study
  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery*
  • Balloon Valvuloplasty* / adverse effects
  • Balloon Valvuloplasty* / mortality
  • Female
  • Germany
  • Heart Valve Prosthesis*
  • Hemodynamics
  • Humans
  • Male
  • Postoperative Complications / mortality
  • Postoperative Complications / therapy
  • Prosthesis Design
  • Prosthesis Failure
  • Recovery of Function
  • Risk Factors
  • Severity of Illness Index
  • Time Factors
  • Transcatheter Aortic Valve Replacement / adverse effects
  • Transcatheter Aortic Valve Replacement / instrumentation*
  • Transcatheter Aortic Valve Replacement / mortality
  • Treatment Outcome

Associated data

  • ClinicalTrials.gov/NCT01645202