Reduced radiation exposure in the cardiac catheterization laboratory with a novel vertical radiation shield

Catheter Cardiovasc Interv. 2020 Jan;95(1):7-12. doi: 10.1002/ccd.28629. Epub 2019 Dec 3.

Abstract

Objectives: Investigation of novel vertical radiation shield (VRS) in reducing operator radiation exposure.

Background: Radiation exposure to the operator remains an occupational health hazard in the cardiac catheterization laboratory (CCL).

Methods: A mannequin simulating an operator was placed near a computational phantom, simulating a patient. Measurement of dose equivalent and Air Kerma located the angle with the highest radiation, followed by a common magnification (8 in.) and comparison of horizontal radiation absorbing pads (HRAP) with or without VRS with two different: CCL, phantoms, and dosimeters. Physician exposure was subsequently measured prospectively with or without VRS during clinical procedures.

Results: Dose equivalent and Air Kerma to the mannequin was highest at left anterior oblique (LAO)-caudal angle (p < .005). Eight-inch magnification increased mGray by 86.5% and μSv/min by 12.2% compared to 10-in. (p < .005). Moving 40 cm from the access site lowered μSv/min by 30% (p < .005). With LAO-caudal angle and 8-in. magnification, VRS reduced μSv/min by 59%, (p < .005) in one CCL and μSv by 100% (p = .016) in second CCL in addition to HRAP. Prospective study of 177 procedures with HRAP, found VRS lowered μSv by 41.9% (μSv: 15.2 ± 13.4 vs. 26.2 ± 31.4, p = .001) with no difference in mGray. The difference was significant after multivariate adjustment for specified variables (p < .001).

Conclusions: Operator radiation exposure is significantly reduced utilizing a novel VRS, HRAP, and distance from the X-ray tube, and consideration of lower magnification and avoiding LAO-caudal angles to lower radiation for both operator and patient.

Keywords: Air Kerma; distance; dose equivalent; horizontal radiation absorbing pad; magnification.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Cardiac Catheterization* / adverse effects
  • Equipment Design
  • Female
  • Humans
  • Lead
  • Male
  • Manikins
  • Middle Aged
  • Occupational Exposure / adverse effects
  • Occupational Exposure / prevention & control*
  • Prospective Studies
  • Protective Devices*
  • Radiation Dosage*
  • Radiation Exposure / adverse effects
  • Radiation Exposure / prevention & control*
  • Radiation Protection / instrumentation*
  • Radiography, Interventional* / adverse effects
  • Risk Assessment
  • Risk Factors
  • Scattering, Radiation

Substances

  • Lead