Background and aims: Multidrug-resistant infectious outbreaks associated with duodenoscope reuse have been documented internationally. A single-use endoscope could eliminate exogenous patient-to-patient infection associated with ERCP.
Methods: We conducted a comparative bench simulation study of a new single-use and 3 models of reusable duodenoscopes on a synthetic anatomic bench model. Four ERCP tasks were performed: guidewire locking (single-use and 1 reusable duodenoscope only), plastic stent placement and removal, metal stent placement and removal, and basket sweeping. The study schedule included block randomization by 4 duodenoscopes, 4 tasks, and 2 anatomic model ERCP stations. Ability to complete tasks, task completion times, and subjective ratings of overall performance, navigation/pushability, tip control, and image quality on a scale of 1 (worst) to 10 (best) were compared among duodenoscopes.
Results: All 4 ERCP tasks (total 14 subtasks) were completed by 6 expert endoscopists using all 4 duodenoscopes, with similar task completion times (median, 1.5-8.0 minutes per task) and overall performance ratings by task (median, 8.0-10.0). Navigation/pushability ratings were lower for the single-use duodenoscope than for the 3 reusable duodenoscopes (median, 8.0, 10.0, 9.0, and 9.0, respectively; P < .01). Tip control ratings were similar among all the duodenoscopes (median, 9.0-10.0; P = .77). Image quality ratings were lower for 1 reusable duodenoscope compared with the single-use and other 2 reusable duodenoscopes (median, 8.0, 9.0, 9.0, and 9.0, respectively; P < .01).
Conclusions: A new single-use duodenoscope was used to simulate 4 ERCP tasks in an anatomic model, with performance ratings and completion times comparable with 3 models of reusable duodenoscopes.
Copyright © 2020 American Society for Gastrointestinal Endoscopy. Published by Elsevier Inc. All rights reserved.