Objective: To evaluate the accuracy of the InBody BP170 oscillometric upper-arm home blood pressure (BP) monitor in general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018).
Methods: Subjects were recruited to fulfill the age, gender, BP and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in general population using the same arm sequential BP measurement method. Two cuffs of the test device were used for arm circumference 22-32 (medium) and 32-42 cm (large).
Results: One-hundred five subjects were recruited and 88 were analyzed. For validation Criterion 1, the mean ± SD of the differences between the test device and reference BP readings was 0.9 ± 6.9/0.5 ± 5.5 mmHg (systolic/diastolic). For Criterion 2, the SD of the averaged BP differences between the test device and reference BP per subject was 5.77/4.80 mmHg (systolic/diastolic).
Conclusion: The InBody BP170 oscillometric home BP monitor fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in general population and can be recommended for clinical use.