Objective: The accuracy of stereoelectroencephalography (SEEG) electrode implantation is an important factor in maximizing its safety. The authors established a quality assurance (QA) process to aid advances in implantation accuracy.
Methods: The accuracy of three consecutive modifications of a frameless implantation technique was quantified in three cohorts comprising 22, 8, and 23 consecutive patients. The modifications of the technique aimed to increase accuracy of the bolt placement.
Results: The lateral shift of the axis of the implanted bolt at the level of the planned entry point was reduced from a mean of 3.0 ± 1.6 mm to 1.4 ± 0.8 mm. The lateral shift of the axis of the implanted bolt at the level of the planned target point was reduced from a mean of 3.8 ± 2.5 mm to 1.6 ± 0.9 mm.
Conclusions: This QA framework helped to isolate and quantify the factors introducing inaccuracy in SEEG implantation, and to monitor ongoing accuracy and the effect of technique modifications.
Keywords: SEEG; accuracy; epilepsy; implantation; quality assurance.