This article describes the design and presents recent results from testing and calibration of a forward Compton scattering high energy X-ray spectrometer. The calibration was performed using a bremsstrahlung source on the photon scattering facility at the Electron linac for beams with high brilliance and low emittance accelerator at Helmholtz-Zentrum Dresden-Rossendorf, which provides high energy X-ray photons with energies up to 18 MeV. The calibration was conducted at different bremsstrahlung end point energies-10.5, 13, 15, and 18 MeV. Experimental spectra show a systematic increase in the maximum energy, photon temperature, and flux. The spectrometer is effective for an energy range of 4-20 MeV with energy resolution. The spectrometer operates in low vacuum with pressure less than 0.1 mbar. Experimental tests showed that operating such a spectrometer in air causes a spuriously enhanced high energy signal due to Compton scattering of photons within air. The article also describes the design and shielding considerations which helped to achieve a dynamic range greater than 30 with this spectrometer. The comparison between the experimental results and Monte Carlo simulations are also presented.