A variable stiffness mechanism for steerable percutaneous instruments: integration in a needle

Med Biol Eng Comput. 2018 Dec;56(12):2185-2199. doi: 10.1007/s11517-018-1847-7. Epub 2018 Jun 4.

Abstract

Needles are advanced tools commonly used in minimally invasive medical procedures. The accurate manoeuvrability of flexible needles through soft tissues is strongly determined by variations in tissue stiffness, which affects the needle-tissue interaction and thus causes needle deflection. This work presents a variable stiffness mechanism for percutaneous needles capable of compensating for variations in tissue stiffness and undesirable trajectory changes. It is composed of compliant segments and rigid plates alternately connected in series and longitudinally crossed by four cables. The tensioning of the cables allows the omnidirectional steering of the tip and the stiffness tuning of the needle. The mechanism was tested separately under different working conditions, demonstrating a capability to exert up to 3.6 N. Afterwards, the mechanism was integrated into a needle, and the overall device was tested in gelatine phantoms simulating the stiffness of biological tissues. The needle demonstrated the capability to vary deflection (from 11.6 to 4.4 mm) and adapt to the inhomogeneity of the phantoms (from 21 to 80 kPa) depending on the activation of the variable stiffness mechanism. Graphical abstract ᅟ.

Keywords: Minimally invasive instruments; Needle deflection; Steerable needle; Tissue inhomogeneity; Variable stiffness.

MeSH terms

  • Equipment Design
  • Gelatin
  • Models, Theoretical*
  • Needles*
  • Phantoms, Imaging

Substances

  • Gelatin