Automatic replanning of VMAT plans for different treatment machines: A template-based approach using constrained optimization

Strahlenther Onkol. 2018 Oct;194(10):921-928. doi: 10.1007/s00066-018-1319-x. Epub 2018 May 30.

Abstract

Purpose: To investigate a new automatic template-based replanning approach combined with constrained optimization, which may be highly useful for a rapid plan transfer for planned or unplanned machine breakdowns. This approach was tested for prostate cancer (PC) and head-and-neck cancer (HNC) cases.

Methods: The constraints of a previously optimized volumetric modulated arc therapy (VMAT) plan were used as a template for automatic plan reoptimization for different accelerator head models. All plans were generated using the treatment planning system (TPS) Hyperion. Automatic replanning was performed for 16 PC cases, initially planned for MLC1 (4 mm MLC) and reoptimized for MLC2 (5 mm) and MLC3 (10 mm) and for 19 HNC cases, replanned from MLC2 to MLC3. EUD, Dmean, D2%, and D98% were evaluated for targets; for OARs EUD and D2% were analyzed. Replanning was considered successful if both plans fulfilled equal constraints.

Results: All prostate cases were successfully replanned. The mean relative target EUD deviation was -0.15% and -0.57% for replanning to MLC2 and MLC3, respectively. OAR sparing was successful in all cases. Replanning of HNC cases from MLC2 to MLC3 was successful in 16/19 patients with a mean decrease of -0.64% in PTV60 EUD. In three cases target doses were substantially decreased by up to -2.58% (PTV60) and -3.44% (PTV54), respectively. Nevertheless, OAR sparing was always achieved as planned.

Conclusions: Automatic replanning of VMAT plans for a different treatment machine by using pre-existing constraints as a template for a reoptimization is feasible and successful in terms of equal constraints.

Keywords: Automatic re-planning; Constrained optimization; Machine failure concept; Radiotherapy; Volumetric modulated arc therapy.

MeSH terms

  • Equipment Design
  • Equipment Failure
  • Humans
  • Male
  • Otorhinolaryngologic Neoplasms / radiotherapy*
  • Parotid Gland / radiation effects
  • Prostatic Neoplasms / radiotherapy*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods*
  • Radiotherapy, Intensity-Modulated / instrumentation*
  • Radiotherapy, Intensity-Modulated / methods*
  • Rectum / radiation effects
  • Spinal Cord / radiation effects
  • Urinary Bladder / radiation effects