Introduction: We studied whether fully automated Elecsys cerebrospinal fluid (CSF) immunoassay results were concordant with positron emission tomography (PET) and predicted clinical progression, even with cutoffs established in an independent cohort.
Methods: Cutoffs for Elecsys amyloid-β1-42 (Aβ), total tau/Aβ(1-42), and phosphorylated tau/Aβ(1-42) were defined against [18F]flutemetamol PET in Swedish BioFINDER (n = 277) and validated against [18F]florbetapir PET in Alzheimer's Disease Neuroimaging Initiative (n = 646). Clinical progression in patients with mild cognitive impairment (n = 619) was studied.
Results: CSF total tau/Aβ(1-42) and phosphorylated tau/Aβ(1-42) ratios were highly concordant with PET classification in BioFINDER (overall percent agreement: 90%; area under the curve: 94%). The CSF biomarker statuses established by predefined cutoffs were highly concordant with PET classification in Alzheimer's Disease Neuroimaging Initiative (overall percent agreement: 89%-90%; area under the curves: 96%) and predicted greater 2-year clinical decline in patients with mild cognitive impairment. Strikingly, tau/Aβ ratios were as accurate as semiquantitative PET image assessment in predicting visual read-based outcomes.
Discussion: Elecsys CSF biomarker assays may provide reliable alternatives to PET in Alzheimer's disease diagnosis.
Keywords: Amyloid PET concordance; Amyloid-β (1–42); Biomarker validation; CSF biomarkers; Clinical progression; Cutoffs; Phosphorylated tau (pTau); Total tau (tTau).