Outcomes for the Commercial Use of Self-Expanding Prostheses in Transcatheter Aortic Valve Replacement: A Report From the STS/ACC TVT Registry

JACC Cardiovasc Interv. 2017 Oct 23;10(20):2090-2098. doi: 10.1016/j.jcin.2017.07.027.

Abstract

Objectives: The authors sought to compare the outcomes of commercial transcatheter aortic valve replacement (TAVR) with the repositionable Evolut R platform to those observed with the CoreValve device in the Society of Thoracic Surgeons (STS)/American College of Cardiology (ACC) Transcatheter Valve Therapy (TVT) Registry.

Background: TAVR continues to evolve, with rapid adoption of iterative changes for commercial practice. Insight into the outcomes of this adoption is needed.

Methods: Patients in the TVT Registry who had TAVR using a 23-, 26-, or 29-mm self-expanding prosthesis were enrolled. Site-reported events for procedural, in-hospital, and 30-day outcomes were examined.

Results: Between January 2014 and April 2016, 9,616 patients underwent TAVR with a self-expanding prosthesis with data entered in the TVT Registry. Compared with patients treated with CoreValve TAVR, those who received Evolut R TAVR had a lower STS-PROM score (8.0 ± 5.4% vs. 8.7 ± 5.3%; p < 0.001), more iliofemoral access (91.6% vs. 89.2%; p < 0.001), and more frequently had conscious sedation (27.4% vs. 12.7%; p < 0.001). With Evolut R TAVR, there was less need for a second prosthesis (2.2% vs. 4.5%; p < 0.001), less device migration (0.2% vs. 0.6%; p = 0.01), a lower incidence of moderate/severe paravalvular regurgitation (post-procedure, 4.4% vs. 6.2%; p < 0.001), and shorter median hospital stay (4.0 vs. 5.0 days; p < 0.001). Patients treated with Evolut R TAVR had greater device success (96.3% vs. 94.9%; p = 0.001). At 30 days, Evolut R patients had both lower mortality (3.7% vs. 5.3%; p < 0.001) and less need for a pacemaker (18.3% vs. 20.1%; p = 0.03).

Conclusions: Commercial adoption of the Evolut R platform is associated with significant improvements in acute outcomes for patients undergoing TAVR for aortic stenosis.

Keywords: CoreValve; Evolut R; aortic stenosis; outcomes; transcatheter.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / physiopathology
  • Aortic Valve / surgery*
  • Aortic Valve Stenosis / diagnostic imaging
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery*
  • Female
  • Heart Valve Prosthesis*
  • Humans
  • Length of Stay
  • Male
  • Postoperative Complications / etiology
  • Postoperative Complications / therapy
  • Prosthesis Design
  • Registries
  • Time Factors
  • Transcatheter Aortic Valve Replacement / adverse effects
  • Transcatheter Aortic Valve Replacement / instrumentation*
  • Transcatheter Aortic Valve Replacement / mortality
  • Treatment Outcome
  • United States