Matched Comparison of Self-Expanding Transcatheter Heart Valves for the Treatment of Failed Aortic Surgical Bioprosthesis: Insights From the Valve-in-Valve International Data Registry (VIVID)

Circ Cardiovasc Interv. 2017 Apr;10(4):e004392. doi: 10.1161/CIRCINTERVENTIONS.116.004392.

Abstract

Background: Transcatheter valve-in-valve implantation is an established therapy for high-risk patients with failed surgical aortic bioprosthesis. There are limited data comparing outcomes of valve-in-valve implantation using different transcatheter heart valves (THV).

Methods and results: Patients included in the Valve-in-Valve International Data registry (VIVID) and treated with self-expanding THV devices were analyzed using centralized core laboratory blinded to clinical events. St. Jude Medical Portico versus Medtronic CoreValve were compared in a 1:2 fashion after propensity score matching. A total of 162 patients, Portico- (n=54) and CoreValve- (n=108) based valve-in-valve procedures comprised the study population with no significant difference in baseline characteristics (age, 79±8.2 years; 60% women; mean STS [Society of Thoracic Surgery] score 8.1±5.5%). Postimplantation, CoreValve was associated with a larger effective orifice area (1.67 versus 1.31 cm2; P=0.001), lower mean gradient (14±7.5 versus 17±7.5 mm Hg; P=0.02), and lower core laboratory-adjudicated moderate-to-severe aortic insufficiency (4.2% versus 13.7%; P=0.04), compared with Portico. Procedural complications including THV malpositioning, second THV requirement, or coronary obstruction were not significantly different between the 2 groups. Survival and stroke rates at 30 days were similar, but overall mortality at 1 year was higher among patients treated with Portico compared with CoreValve (22.6% versus 9.1%; P=0.03).

Conclusions: In this first matched comparison of THVs for valve-in-valve implantations, Portico and CoreValve demonstrated differences in postprocedural hemodynamics and long-term clinical outcomes. Although this could be related to THV design characteristics, the impact of other procedural factors cannot be excluded and require further evaluation.

Keywords: aortic valve; bioprosthesis; hemodynamic; transcatheter aortic valve replacement.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Aged
  • Aortic Valve / surgery*
  • Aortic Valve Stenosis / physiopathology
  • Aortic Valve Stenosis / surgery*
  • Bioprosthesis / adverse effects*
  • Europe / epidemiology
  • Female
  • Follow-Up Studies
  • Heart Valve Prosthesis / adverse effects*
  • Hemodynamics
  • Humans
  • Male
  • Middle East / epidemiology
  • New Zealand / epidemiology
  • Propensity Score
  • Prospective Studies
  • Prosthesis Design
  • Registries*
  • Reoperation / methods
  • Survival Rate / trends
  • Transcatheter Aortic Valve Replacement / methods*
  • Transcatheter Aortic Valve Replacement / mortality
  • Treatment Outcome
  • United States / epidemiology