Time-multiplexed structured illumination using a DMD for optical diffraction tomography

Opt Lett. 2017 Mar 1;42(5):999-1002. doi: 10.1364/OL.42.000999.

Abstract

We present a time-multiplexing structured illumination control technique for optical diffraction tomography (ODT). Instead of tilting the angle of illumination, time-multiplexed sinusoidal illumination is exploited using a digital micromirror device (DMD). The present method effectively eliminates unwanted diffracted beams from binary DMD patterns, which deteriorates the image quality of the ODT in the previous binary Lee hologram method. We experimentally show the feasibility and advantage of the present method by reconstructing three-dimensional refractive index distributions of various samples and comparing with a conventional Lee hologram method.