Left ventricular assist device (LVAD) implantation can be challenging in patients with a prior surgical ventricular restoration (SVR). In this case series of heart failure patients with a history of SVR, we describe the surgical technique and outcome of a customized approach for inflow cannula orientation. Seven patients with a history of SVR with end-stage chronic heart failure were accepted for long-term LVAD support. In all patients, the Dacron patch was removed through left ventriculotomy and a Hegar 22 dilator was inserted at the estimated optimal position of the LVAD inflow cannula. The left ventricle was reconstructed around the dilator from the left ventricular (LV) apex to the base. Finally, the LVAD sewing ring was sutured onto the remaining apical defect and a HeartWare® LVAD was implanted. LVAD implantation was successful in all 7 patients. Transoesophageal echocardiography ensured an adequate LVAD position and inflow and outflow cannula Doppler flow recordings. The mean intensive care unit stay was 5.8 ± 2.6 days, and the hospital stay after surgery was 32 ± 16 days. All patients follow regular visits (follow-up 20 ± 16 months) at the outpatient clinic without any remarkable event. Using the technique described, LVAD implantation in patients after SVR is feasible and safe.
Keywords: Heart failure; Left ventricular assist device; Surgical ventricular reconstruction.
© The Author 2016. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.