A rapid liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed and optimized for the simultaneous determination of bisphenol A, 4-t-octylphenol and 4-nonylphenol in human blood serum. For the first time, the electrospray ionization (ESI) parameters of probe position, voltage potential, sheath gas flow rate, auxiliary gas flow rate, and ion transfer tube temperature were thoroughly studied and optimized for each phenol by a univariate approach. As a consequence, low instrumental limits of detection were reported, demonstrating at 0.2 ng/mL (in solvent matrix) excellent injection repeatability (RSD<14.5%) and a confirmation peak for all target phenols. Extraction and purification of serum was performed by the novel Hybrid Solid Phase Extraction-Precipitation Technology technique (Hybrid SPE-PPT). The limits of detection in human blood serum were 0.80, 1.3 and 1.4 ng/mL for BPA, 4-t-OP and 4-NP, respectively.
Keywords: 4-Nonylphenol; 4-t-Octylphenol; Bioanalysis; Bisphenol A; EDCs; Hybrid SPE-PPT.
Copyright © 2015 Elsevier B.V. All rights reserved.