Achieving safe hands-on defibrillation using electrical safety gloves--a clinical evaluation

Resuscitation. 2015 May:90:163-7. doi: 10.1016/j.resuscitation.2014.12.028. Epub 2015 Feb 26.

Abstract

Introduction: Safe hands-on defibrillation (HOD) will allow uninterrupted chest compression during defibrillation and may improve resuscitation success. We tested the ability of electrical insulating gloves to protect the rescuer during HOD using a 'worst case' electrical scenario.

Materials and method: Leakage current flowing from the patient to the 'rescuer' during antero-lateral defibrillation of patients undergoing elective cardioversion was measured. The 'rescuer' maintained firm (20 kgf) contact with the patient during defibrillation, wearing Class 1 electrical insulating gloves while simulating an inadvertent contact with the patient, through an additional wired contact between 'rescuer' and patient.

Results: Data from 61 shocks from 43 different patients were recorded. The median leakage current from all defibrillations was 20.0 μA, (range: 2.0-38.5). In total, 18 of the shocks were delivered at 360 J and had a median leakage current of 27.0 μA (range: 14.3-38.5).

Conclusion: When using Class 1 electrical insulating gloves for hands-on defibrillation, rescuer leakage current is significantly below the 1 mA safe threshold, allowing safe hands-on defibrillation if the rescuer makes only one other point of contact with the patient.

Keywords: Defibrillation; Outcome; Resuscitation; Safety.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Defibrillators*
  • Electric Countershock*
  • Electric Injuries / prevention & control*
  • Feasibility Studies
  • Gloves, Protective*
  • Heart Arrest / therapy*
  • Humans
  • Middle Aged
  • Occupational Injuries / prevention & control*