Aim: Preclinical studies have suggested that nitinol-based compression anastomosis might be a viable solution to anastomotic leak following low anterior resection. A prospective multicentre open label study was therefore designed to evaluate the performance of the ColonRing(™) in (low) colorectal anastomosis.
Method: The primary outcome measure was anastomotic leakage. Patients were recruited at 13 different colorectal surgical units in Europe, the United States and Israel. Institutional review board approval was obtained.
Results: Between 21 March 2010 and 3 August 2011, 266 patients completed the study protocol. The overall anastomotic leakage rate was 5.3% for all anastomoses, including a rate of 3.1% for low anastomoses. Septic anastomotic complications occurred in 8.3% of all anastomoses and 8.2% of low anastomoses.
Conclusion: Nitinol compression anastomosis is safe, effective and easy to use and may offer an advantage for low colorectal anastomosis. A prospective randomized trial comparing ColonRing(™) with conventional stapling is needed.
Keywords: Compression anastomosis; anastomotic leakage; low anterior resection.
Colorectal Disease © 2014 The Association of Coloproctology of Great Britain and Ireland.