Incidence, predictors, and procedural results of upgrade to resynchronization therapy: the RAFT upgrade substudy

Circ Arrhythm Electrophysiol. 2015 Feb;8(1):152-8. doi: 10.1161/CIRCEP.114.001997. Epub 2014 Nov 23.

Abstract

Background: The resynchronization-defibrillation for ambulatory heart failure trial (RAFT) study demonstrated that adding cardiac resynchronization therapy (CRT) in selected patients requiring de novo implantable cardiac defibrillators (ICD) reduced mortality as compared with ICD therapy alone, despite an increase in procedure-related adverse events. Data are lacking regarding the management of patients with ICD therapy who develop an indication for CRT upgrade.

Methods and results: Participating RAFT centers provided data regarding de novo CRT-D (CRT with ICD) implant, upgrade to CRT-D during RAFT (study upgrade), and upgrade within 6 months after presentation of study results (substudy). Substudy centers enrolled 1346 (74.9%) patients in RAFT, including 644 de novo, 80 study upgrade, and 60 substudy CRT attempts. The success rate (initial plus repeat attempts) was 95.2% for de novo versus 96.3% for study upgrade and 90.0% for substudy CRT attempts (P=0.402). Acute complications occurred among 26.2% of de novo versus 18.8% of study upgrade and 3.4% of substudy CRT implantation attempts (P<0.001). The most common complication was left ventricular lead dislodgement. The principal reasons for not yet attempting upgrade in the substudy were patient preference (31.9%), New York Heart Association Class I (17.0%), and a QRS<150 ms (13.1%).

Conclusions: Among a broad group of implant physicians, CRT upgrades were performed in patients with an ICD in situ with no difference in implant success rate and a reduced acute complication rate as compared with a de novo CRT implant. Decisions to upgrade were influenced by predictors of benefit in subgroup analyses of the RAFT study and other trials.

Keywords: cardiac resynchronization therapy; complication; implantable cardioverter-defibrillator; implantation.

Publication types

  • Multicenter Study
  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Cardiac Resynchronization Therapy Devices
  • Cardiac Resynchronization Therapy* / adverse effects
  • Cardiac Resynchronization Therapy* / mortality
  • Defibrillators, Implantable
  • Electric Countershock* / instrumentation
  • Female
  • Heart Failure / diagnosis
  • Heart Failure / mortality
  • Heart Failure / physiopathology
  • Heart Failure / therapy*
  • Humans
  • Male
  • Middle Aged
  • Patient Preference
  • Patient Selection
  • Risk Factors
  • Time Factors
  • Treatment Outcome
  • United States
  • Ventricular Dysfunction, Left / diagnosis
  • Ventricular Dysfunction, Left / mortality
  • Ventricular Dysfunction, Left / physiopathology
  • Ventricular Dysfunction, Left / therapy*