Mechanical circulatory support of a univentricular Fontan circulation with a continuous axial-flow pump in a piglet model

ASAIO J. 2015 Mar-Apr;61(2):196-201. doi: 10.1097/MAT.0000000000000172.

Abstract

Despite the significant contribution of the Fontan procedure to the therapy of complex congenital heart diseases, many patients progress to failure of their Fontan circulation. The use of ventricular assist devices to provide circulatory support to these patients remains challenging. In the current study, a continuous axial-flow pump was used to support a univentricular Fontan circulation. A modified Fontan circulation (atrio-pulmonary connection) was constructed in six Yorkshire piglets (8-14 kg). A Dacron conduit (12 mm) with two branches was constructed to serve as a complete atrio-pulmonary connection without the use of cardiopulmonary bypass. The Impella pump was inserted into the conduit through an additional Polytetrafluoroethylene (PTFE) graft in five animals. Hemodynamic data were collected for 6 hours under the supported Fontan circulation. The control animal died after initiating the Fontan circulation independent of resuscitation. Four pump supported animals remained hemodynamically stable for 6 hours with pump speeds between 18,000 rpm and 22,000 rpm (P1-P3). Oxygen saturation was maintained between 95% and 100%. Normal organ perfusion was illustrated by blood gas analysis and biochemical assays. A continuous axial-flow pump can be used for temporal circulatory support to the failing Fontan circulation as "bridge" to heart transplantation or recovery.

MeSH terms

  • Animals
  • Assisted Circulation / instrumentation*
  • Assisted Circulation / methods
  • Fontan Procedure / adverse effects
  • Fontan Procedure / instrumentation*
  • Fontan Procedure / methods
  • Heart-Assist Devices
  • Hemodynamics
  • Lactic Acid / blood
  • Miniaturization
  • Models, Animal
  • Oxygen / blood
  • Sus scrofa

Substances

  • Lactic Acid
  • Oxygen