Purpose: The need for rapid and accurate H typing is evident during Escherichia coli outbreak situations. This study explores the transition of MS-H, a method originally developed for rapid H antigen typing of E. coli using LC-MS/MS of flagella digest of reference strains and some clinical strains, to E. coli isolates in clinical scenario through quantitative analysis and method validation.
Experimental design: Motile and nonmotile strains were examined in batches to simulate clinical sample scenario. Various LC-MS/MS batch run procedures and MS-H typing rules were compared and summarized through quantitative analysis of MS-H data output for a standard method development.
Results: Label-free quantitative data analysis of MS-H typing was proven very useful for examining the quality of MS-H result and the effects of some sample carryovers from motile E. coli isolates. Based on this, a refined procedure and protein identification rule specific for clinical MS-H typing was established and validated.
Conclusions and clinical relevance: With LC-MS/MS batch run procedure and database search parameter unique for E. coli MS-H typing, the standard procedure maintained high accuracy and specificity in clinical situations, and its potential to be used in a clinical setting was clearly established.
Keywords: E. coli; Flagella typing; LC-MS/MS; Mass spectrometry; Quality control.