Comparison of the ultra-low-dose Veo algorithm with the gold standard filtered back projection for detecting pulmonary asbestos-related conditions: a clinical observational study

BMJ Open. 2014 May 30;4(5):e004980. doi: 10.1136/bmjopen-2014-004980.

Abstract

Objectives: Radiation delivered during CT is a major concern, especially for individuals undergoing repeated screening. We aimed to compare a new ultra-low-dose algorithm called Veo with the gold standard filtered back projection (FBP) for detecting pulmonary asbestos-related conditions.

Setting: University Hospital CHU G. Montpied, Clermont-Ferrand, France

Participants: Asbestos-exposed workers were recruited following referral to screening for asbestos-related conditions. Two acquisitions were performed on a 64-slice CT: the gold standard FBP followed by Veo reconstruction.

Outcome measures: Two radiologists independently assessed asbestos-related abnormalities, pulmonary nodules, radiation doses and image quality (noise).

Results: We included 27 asbestos-exposed workers (63.3±6.5 years with 11.9±9.7 years of asbestos exposure). We observed 297 pleural plaques in 20 participants (74%). All patients (100%) had pulmonary nodules, totalling 167 nodules. Detection rates did not differ for pleural plaques (Veo 87% vs FBP 97%, NS), pleural thickening (100% for both) and pulmonary nodules (80% for both). Interstitial abnormalities were depicted less frequently with Veo than FBP. False negative and false positive did not exceed 2.7%. Compared with FBP, Veo decreased the radiation dose up to 87% (Veo 0.23±0.07 vs FBP 1.83±0.88 mSv, p<0.001). The objective image noise also decreased with Veo as much as 23% and signal-to-noise ratio increased up to 33%.

Conclusions: A low-dose CT with Veo reconstruction substantially reduced radiation. Veo compared favourably with FBP in detecting pleural plaques, pleural thickening and pulmonary nodules. These results should be confirmed on a larger sample size before the use of Veo in clinical routine practice in asbestos-related conditions, especially regarding the low prevalence of interstitial abnormalities in this study.

Trial registration number: NCT01955018.

Keywords: OCCUPATIONAL & INDUSTRIAL MEDICINE; PUBLIC HEALTH.

Publication types

  • Comparative Study
  • Observational Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms*
  • Asbestosis / diagnostic imaging*
  • Humans
  • Middle Aged
  • Radiation Dosage*
  • Radiation Exposure / prevention & control*
  • Radiation Exposure / standards
  • Radiography

Associated data

  • ClinicalTrials.gov/NCT01955018