Use of the Nexfin™ device to detect acute arterial pressure variations during anaesthesia induction

Br J Anaesth. 2014 Jul;113(1):52-60. doi: 10.1093/bja/aeu055. Epub 2014 Apr 25.

Abstract

Background: Standard non-invasive arterial pressure (AP) measurements are discontinuous. By providing non-invasive beat-to-beat AP measurements, Nexfin™ might limit duration of intraoperative hypotension and hypertension. We assessed the ability of Nexfin™ to detect AP variations by comparing its trending ability with invasive AP monitoring.

Methods: Thirty-one subjects undergoing elective surgery under general anaesthesia were included. During induction, simultaneous pairs of AP measurements were collected every 5 s from the Nexfin™ finger sensor and a homolateral radial artery catheter. Magnitude and time lags of AP variations from baseline to nadir and peak were calculated for both methods. Concordance analysis was performed by the Bland-Altman method (for comparison of repeated measures when appropriate).

Results: Nexfin™ detected 100% of AP changes with the median delays of 0 s (-13 to 7) and 0 s (-5 to 12) for nadir and peak, respectively. Bias [limits of agreement (LOA)] of systolic AP (SAP) variations was -0.5 mm Hg (-31.2 to 30.2) and -9.4 mm Hg (-31.3 to 12.6) from baseline to nadir and from baseline to peak, respectively. For 3479 analysed paired measurements, bias was -3.8 and -8.8 mm Hg for SAP and diastolic AP, with LOA of (-36.0 to 28.5) and (-29.8 to 12.3), respectively.

Conclusions: Nexfin™ detects AP variations accurately and can be a useful warning device during anaesthesia. However, it is not interchangeable with invasive monitoring, given the large LOA between the two measurements.

Clinical trial registration: NCT01658631.

Keywords: anaesthesiology; arterial pressure monitors; monitoring, intraoperative.

Publication types

  • Clinical Trial

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Anesthesia, General / methods*
  • Blood Pressure / physiology
  • Blood Pressure Determination / instrumentation
  • Blood Pressure Determination / methods
  • Blood Pressure Monitors*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Monitoring, Intraoperative / instrumentation*
  • Monitoring, Intraoperative / methods
  • Prospective Studies
  • Radial Artery / physiology
  • Reproducibility of Results
  • Young Adult

Associated data

  • ClinicalTrials.gov/NCT01658631