Abstract
In an effort to enhance the linear range of anti-CCP we developed a new immunoassay based on time-resolved fluoroimmunoassay. The precision, sensitivity, specificity, and stability of the assay were evaluated ELISA set as control. The anti-CCP IgG TRFIA kit we established had a wider detectable range than commercial ELISA ones. With regard to intra- and inter-assay precision, the TRFIA kit was better than threee commercial ELISA ones. The mean recovery rate was 101.0%. The TRFIA we developed for anti-CCP IgG detection yielded a more sensitive and reliable method for RA diagnosis and large-scale screening programs as well.