Image quality with single-heartbeat 320-multidetector computed tomographic coronary angiography

J Comput Assist Tomogr. 2014 May-Jun;38(3):444-50. doi: 10.1097/RCT.0000000000000044.

Abstract

Objective: We aimed to establish the feasibility of single-heartbeat 320-multidetector computed tomographic coronary angiography (CTCA) and assess variables affecting image quality.

Methods: Consecutive patients (n = 249, 38% male) underwent CTCA. Two observers assessed image quality using a 4-point scale (1, excellent; 4, poor).

Results: Mean heart rate was 60 beats per minute (95% confidence interval, 59-62); body mass index, 29 kg/m (28-30); and dose-length product, 283 mGy·cm (266-301). During scanning, 133 (51%) received sublingual glyceryl trinitrate (GTN), 9 (4%) had ectopics, and 12 (5%) had atrial fibrillation. Diagnostic image quality was obtained in 99% with mean image quality of 1.4 (1.3, 1.5). Age, sex, atrial fibrillation, ectopics, diabetes mellitus (12%), and obstructive disease were not related to image quality. A lower heart rate and GTN were associated with improved image quality (P ≤ 0.001).

Conclusions: Optimal image quality in single-heartbeat 320-multidetector CTCA is achievable in 99% of unselected patients. Image quality is improved by lower heart rate and GTN.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Algorithms
  • Cardiac-Gated Imaging Techniques / methods*
  • Coronary Angiography / methods*
  • Coronary Artery Disease / diagnostic imaging*
  • Coronary Artery Disease / physiopathology*
  • Feasibility Studies
  • Female
  • Heart Rate
  • Humans
  • Male
  • Middle Aged
  • Multidetector Computed Tomography / methods*
  • Observer Variation
  • Radiographic Image Enhancement / methods*
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity