Stereoscopic digital mammography: improved specificity and reduced rate of recall in a prospective clinical trial

Radiology. 2013 Jan;266(1):81-8. doi: 10.1148/radiol.12120382. Epub 2012 Nov 13.

Abstract

Purpose: To compare stereoscopic digital mammography (DM) with standard DM for the rate of patient recall and the detection of cancer in a screening population at elevated risk for breast cancer.

Materials and methods: Starting in September 2004 and ending in December 2007, this prospective HIPAA-compliant, institutional review board-approved screening trial, with written informed consent, recruited female patients at elevated risk for breast cancer (eg, personal history of breast cancer or breast cancer in a close relative). A total of 1298 examinations from 779 patients (mean age, 58.6 years; range, 32-91 years) comprised the analyzable data set. A paired study design was used, with each enrolled patient serving as her own control. Patients underwent both DM and stereoscopic DM examinations in a single visit, findings of which were interpreted independently by two experienced radiologists, each using a Breast Imaging Reporting and Data System (BI-RADS) assessment (BI-RADS category 0, 1, or 2). All patients determined to have one or more findings with either or both modalities were recalled for standard diagnostic evaluation. The results of 1-year follow-up or biopsy were used to determine case truth.

Results: Compared with DM, stereoscopic DM showed significantly higher specificity (91.2% [1167 of 1279] vs 87.8% [1123 of 1279]; P = .0024) and accuracy (90.9% [1180 of 1298] vs 87.4% [1135 of 1298]; P = .0023) for detection of cancer. Sensitivity for detection of cancer was not significantly different for stereoscopic DM (68.4% [13 of 19]) compared with DM (63.2% [12 of 19], P .99). The recall rate for stereoscopic DM was 9.6% (125 of 1298) and that for DM was 12.9% (168 of 1298) (P = .0018).

Conclusion: Compared with DM, stereoscopic DM significantly improved specificity for detection of cancer, while maintaining comparable sensitivity. The recall rate was significantly reduced with stereoscopic DM compared with DM.

Supplemental material: http://radiology.rsna.org/lookup/suppl/doi:10.1148/radiol.12120382/-/DC1.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Breast Neoplasms / diagnostic imaging*
  • Breast Neoplasms / epidemiology*
  • Female
  • Georgia / epidemiology
  • Humans
  • Imaging, Three-Dimensional / statistics & numerical data*
  • Mammography / statistics & numerical data*
  • Middle Aged
  • Prevalence
  • Prospective Studies
  • Radiographic Image Enhancement / methods*
  • Reproducibility of Results
  • Risk Factors
  • Sensitivity and Specificity