Effect of high-energy neutral particles on extreme ultraviolet spectroscopy in large helical device

Rev Sci Instrum. 2012 Oct;83(10):10D509. doi: 10.1063/1.4733733.

Abstract

Spectra measured by an extreme ultraviolet (EUV) spectrometer frequently suffer large spike noise when Large Helical Device is operated in low-density range (≤ 3 × 10(13) cm(-3)) with neutral beam injection (NBI). The spike noise completely disappears in electron cyclotron heating discharges. In order to examine the effect of NBI, a carbon filter with thickness of 150 nm was installed in the EUV spectrometer. As a result, the spike noise was reduced by an order of magnitude. It is experimentally verified that the spike noise is caused by escaping high-energy neutral particles resulting from the circulating high-energy hydrogen ions borne from NBI.