Transcatheter valve-in-valve implantation for failed balloon-expandable transcatheter aortic valves

JACC Cardiovasc Interv. 2012 May;5(5):571-577. doi: 10.1016/j.jcin.2012.03.008.

Abstract

Objectives: This study sought to evaluate outcomes after implantation of a second transcatheter heart valve (THV-in-THV) for acute THV failure.

Background: Aortic regurgitation after transcatheter aortic valve replacement (TAVR) may be valvular due to prosthetic leaflet dysfunction or paravalvular due to poor annular sealing.

Methods: Patients undergoing aortic balloon-expandable TAVR at 3 centers were prospectively evaluated at baseline, intraprocedurally, at hospital discharge, and annually.

Results: Of 760 patients undergoing TAVR, 21 (2.8%) received a THV-in-THV implant due to acute, severe regurgitation. Aortic regurgitation was paravalvular in 18 patients and transvalvular in the remaining 3 patients. THV-in-THV implantation was technically successful in 19 patients (90%) and unsuccessful in 2 patients (10%), who subsequently underwent open heart surgery. Mortality at 30 days and 1 year was 14.3% and 24%, respectively. After successful THV-in-THV, mean aortic valve gradient fell from 37 ± 12 mm Hg to 13 ± 5 mm Hg (p < 0.01); aortic valve area increased from 0.64 ± 0.14 cm(2) to 1.55 ± 0.27 cm(2) (p < 0.01); and paravalvular aortic regurgitation was none in 4 patients, mild in 13 patients, and moderate in 2 patients. At 1-year follow-up, 1 patient had moderate and the others had mild or no paravalvular leaks. The mean transvalvular gradient was 15 ± 4 mm Hg, which was higher than in patients undergoing conventional TAVR (11 ± 4 mm Hg, p = 0.02).

Conclusions: THV-in-THV implantation is feasible and results in satisfactory short- and mid-term outcomes.

Publication types

  • Multicenter Study

MeSH terms

  • Acute Disease
  • Aged
  • Aged, 80 and over
  • Aortic Valve Insufficiency / diagnosis
  • Aortic Valve Insufficiency / etiology
  • Aortic Valve Insufficiency / mortality
  • Aortic Valve Insufficiency / therapy*
  • Aortic Valve Stenosis / mortality
  • Aortic Valve Stenosis / therapy*
  • Canada
  • Cardiac Catheterization / instrumentation*
  • Cardiac Catheterization / mortality
  • Chi-Square Distribution
  • Echocardiography, Doppler, Color
  • Echocardiography, Transesophageal
  • Feasibility Studies
  • Female
  • Heart Valve Prosthesis Implantation / adverse effects
  • Heart Valve Prosthesis Implantation / instrumentation*
  • Heart Valve Prosthesis Implantation / methods*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis*
  • Humans
  • Kaplan-Meier Estimate
  • Male
  • Ohio
  • Predictive Value of Tests
  • Prospective Studies
  • Prosthesis Design
  • Prosthesis Failure*
  • Retreatment
  • Risk Assessment
  • Risk Factors
  • Severity of Illness Index
  • Time Factors
  • Treatment Failure