Clear and complete documentation of the physical parameters of total body irradiation (TBI) is one of the essential requirements for the evaluation and improvement of the clinical results of TBI. Concerning the dosimetric aspects of TBI, a number of recommendations have been formulated with emphasis on basic dosimetry, patient dosimetry and dose specification. The dosimeters should be calibrated regularly with reference to the absorbed dose in water. Depth dose measurements should be performed in water equivalent phantoms of specified dimensions. It has been strongly suggested to measure the absorbed dose at the surface of the patient at 8 different regions at the entry and exit of the beam under TBI conditions. The reference dose to the patient should be specified as the total dose to mid abdomen at the height of the umbilicus. As an independent parameter, the lung dose should be specified as the mean dose in the central region of the shielded part of both lungs. Recent, more complete, information on the physical and dosimetric aspects of TBI will be incorporated in the registry of the European Bone Marrow Transplant Group (EBMT). A cooperation has been established between the EBMT and the European Late Effects Project Group (EULEP) to study the development of late effects in man caused by ionising radiation.