3-dimensional aortic annular assessment by multidetector computed tomography predicts moderate or severe paravalvular regurgitation after transcatheter aortic valve replacement: a multicenter retrospective analysis

J Am Coll Cardiol. 2012 Apr 3;59(14):1287-94. doi: 10.1016/j.jacc.2011.12.015. Epub 2012 Feb 22.

Abstract

Objectives: This study sought to analyze multidetector computed tomography (MDCT) 3-dimensional aortic annular dimensions for the prediction of paravalvular aortic regurgitation (PAR) following transcatheter aortic valve replacement (TAVR).

Background: Moderate or severe PAR after TAVR is associated with increased morbidity and mortality.

Methods: A total of 109 consecutive patients underwent MDCT pre-TAVR with a balloon expandable aortic valve. Differences between transcatheter heart valve (THV) size and MDCT measures of annular size (mean diameter, area, and circumference) were analyzed concerning prediction of PAR. Patients with THV malposition (n = 7) were excluded. In 50 patients, MDCT was repeated after TAVR to assess THV eccentricity (1 - short diameter/long diameter) and expansion (MDCT measured THV area/nominal THV area).

Results: Moderate or severe PAR (13 of 102) was associated with THV undersizing (THV diameter - mean diameter = -0.7 ± 1.4 mm vs. 0.9 ± 1.8 mm for trivial to mild PAR, p < 0.01). The difference between THV size and MDCT annular size was predictive of PAR (mean diameter: area under the curve [AUC]: 0.81, 95% confidence interval [CI]: 0.68 to 0.88; area: AUC: 0.80, 95% CI: 0.65 to 0.90; circumference: AUC: 0.76, 95% CI: 0.59 to 0.91). Annular eccentricity was not associated with PAR (AUC: 0.58, 95% CI: 0.46 to 0.75). We found that 35.3% (36 of 102) and 45.1% (46 of 102) of THVs were undersized relative to the MDCT mean diameter and area, respectively. THV oversizing relative to the annular area was not associated with THV eccentricity or underexpansion (oversized vs. undersized THVs; expansion: 102.7 ± 5.3% vs. 106.1 ± 5.6%, p = 0.03; eccentricity: median: 1.7% [interquartile range: 1.4% to 3.0%] vs. 1.7% [interquartile range: 1.1% to 2.7%], p = 0.28).

Conclusions: MDCT-derived 3-dimensional aortic annular measurements are predictive of moderate or severe PAR following TAVR. Oversizing of THVs may reduce the risk of moderate or severe PAR.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Analysis of Variance
  • Aortic Valve / diagnostic imaging*
  • Aortic Valve / surgery*
  • Aortic Valve Insufficiency / diagnostic imaging*
  • Aortic Valve Insufficiency / etiology
  • Aortic Valve Insufficiency / prevention & control*
  • Cardiac Catheterization / adverse effects
  • Cardiac Catheterization / methods*
  • Cohort Studies
  • Echocardiography, Transesophageal / methods
  • Female
  • Follow-Up Studies
  • Heart Valve Diseases / diagnosis
  • Heart Valve Diseases / surgery
  • Heart Valve Prosthesis Implantation / methods*
  • Humans
  • Imaging, Three-Dimensional
  • Male
  • Minimally Invasive Surgical Procedures / adverse effects
  • Minimally Invasive Surgical Procedures / methods
  • Postoperative Complications / diagnostic imaging*
  • Postoperative Complications / prevention & control
  • Predictive Value of Tests
  • Preoperative Care / methods
  • Retrospective Studies
  • Risk Assessment
  • Severity of Illness Index
  • Statistics, Nonparametric
  • Tomography, X-Ray Computed
  • Treatment Outcome